Dual Channel Flame Photometer
![](http://mantilabsolutions.com/wp-content/uploads/2021/09/Digital-Flame-Photometer-2-min.jpg)
Item Code: MT-126
Dual Channel Flame Photometer
Digital Flame Photometer is a dual channel flame photometer used for simultaneous observation of Sodium, Potassium in blood serum (plasma), Urine & other type of biological fluids.
The instruments are supplied in two units viz. Main Unit and Compressor Unit. The Main Unit comprises of digital displays, detector, filters and nebulizer. The compressor unit supplies the compressed air to the main unit.
Air at a given pressure is passed into the atomiser. The suction produced by it, draws the sample solution into the nebulizer and is sprayed as fine mist into the flame. The light emitted by the element at the characteristic wavelength is isolated by the narrow band interference filters, which then falls on to the silicon photodiode. After signal conditioning of the photodiode output, it is sent to digital converter and the results are displayed directly in ppm/meq units. The narrow band interference filters are mounted on a rotating disc fitted within the Main Unit. Appropriate filter can be selected by rotating this disc with the thumb. The oil free compressor with regulator maintains a stable airflow.
The instruments are supplied in two units viz. Main Unit and Compressor Unit. The Main Unit comprises of digital displays, detector, filters and nebulizer. The compressor unit supplies the compressed air to the main unit.
Air at a given pressure is passed into the atomiser. The suction produced by it, draws the sample solution into the nebulizer and is sprayed as fine mist into the flame. The light emitted by the element at the characteristic wavelength is isolated by the narrow band interference filters, which then falls on to the silicon photodiode. After signal conditioning of the photodiode output, it is sent to digital converter and the results are displayed directly in ppm/meq units. The narrow band interference filters are mounted on a rotating disc fitted within the Main Unit. Appropriate filter can be selected by rotating this disc with the thumb. The oil free compressor with regulator maintains a stable airflow.
Some Key features are:
- Simple Easy to Operate
- Narrow Band Interference Filters
- Accurate Na, K, Ca, Li Determinatio
- Mixing Chamber Designed For Stable Flame
- Concentric Non-Corrosive Nebulizer
Model
MT-126
Range
Serum
Na: 0 – 200 meq/l
K : 0 – 10 meq/l
–
Li: 0 – 2 meq/l (Optional)
Urine
Na: 0 – 200 meq/l
K : 0 – 100 meq/l (Optional)
Ca: 0 – 10 meq/l
–
Sensitivity
–
Accuracy
+ 2% upto 100 meq
+ 5% above 100 meq
Measurement System
2Two elements at a Time
Readout
Dual 2½ Digit, 7-Segment LED
Ignition System
In-built electronic Ignition by press of switch
Repeatability
+ 2 Counts
Detector
Silicon Photodiode
Filters
Narrow band interference glass filters
Nebulizer
Black backelite, axial flow type
Flame System
LPG dry oil free air
Warm up time
10 Minutes
Power
230 V + 10% AC, 50 Hz
Dimensions
375 x 245 x 220 mm (L x B x H) (Approx.)
Weight
7.5 Kg (Approx.)
COMPRESSOR UNIT
Air Supply
By oil free mini compressor unit with pressure regulator
Combustion Gas
LPG controlled by precision regulator
Power
230 V 10% AC, 50 Hz
Dimensions
290 x 255 x 210 mm (L x B x H) (Approx.)
Weight
8 Kg (Approx.)